Non-catching type gauges are the emerging class of in situ precipitation measurement instruments. For these instruments, rigorous testing and calibration are more challenging than for traditional gauges. Hydrometeors characteristics like particle size, shape, fall velocity and density must be reproduced in a controlled environment to provide the reference precipitation, instead of the equivalent water flow used for catching-type gauges. They are generally calibrated by the manufacturers using internal procedures developed for the specific technology employed. No agreed methodology exists, and the adopted procedures are rarely traceable to internationally recognized standards. The EURAMET project 18NRM03 ‘INCIPIT’ on ‘Calibration and accuracy of non-catching instruments to measure liquid/solid atmospheric precipitation’, funded by the European Metrology Programme for Innovation and Research (EMPIR), was initiated in 2019 to investigate calibration and accuracy issues of non-catching measuring instruments used for liquid/solid atmospheric precipitation measurement. A survey of the existing models of non-catching type instruments was initially performed and this paper provides an overview and a description of their working principles and the adopted calibration procedures. Both literature works and technical manuals disclosed by manufacturers are summarized and discussed, while current limitations and metrological requirements are identified.

Calibration of non‐catching precipitation measurement instruments: A review / Lanza, L. G.; Merlone, A.; Cauteruccio, A.; Chinchella, E.; Stagnaro, M.; Dobre, M.; Garcia Izquierdo, M. C.; Nielsen, J.; Kjeldsen, H.; Roulet, Y. A.; Coppa, G.; Musacchio, C.; Bordianu, C.; Parrondo, M.. - In: METEOROLOGICAL APPLICATIONS. - ISSN 1350-4827. - 28:3(2021). [10.1002/met.2002]

Calibration of non‐catching precipitation measurement instruments: A review

Merlone, A.;Coppa, G.;Musacchio, C.
;
2021

Abstract

Non-catching type gauges are the emerging class of in situ precipitation measurement instruments. For these instruments, rigorous testing and calibration are more challenging than for traditional gauges. Hydrometeors characteristics like particle size, shape, fall velocity and density must be reproduced in a controlled environment to provide the reference precipitation, instead of the equivalent water flow used for catching-type gauges. They are generally calibrated by the manufacturers using internal procedures developed for the specific technology employed. No agreed methodology exists, and the adopted procedures are rarely traceable to internationally recognized standards. The EURAMET project 18NRM03 ‘INCIPIT’ on ‘Calibration and accuracy of non-catching instruments to measure liquid/solid atmospheric precipitation’, funded by the European Metrology Programme for Innovation and Research (EMPIR), was initiated in 2019 to investigate calibration and accuracy issues of non-catching measuring instruments used for liquid/solid atmospheric precipitation measurement. A survey of the existing models of non-catching type instruments was initially performed and this paper provides an overview and a description of their working principles and the adopted calibration procedures. Both literature works and technical manuals disclosed by manufacturers are summarized and discussed, while current limitations and metrological requirements are identified.
File in questo prodotto:
File Dimensione Formato  
met.2002_compressed.pdf

accesso aperto

Descrizione: Articolo principale
Tipologia: final published article (publisher’s version)
Licenza: Creative Commons
Dimensione 656.04 kB
Formato Adobe PDF
656.04 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/71312
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 21
  • ???jsp.display-item.citation.isi??? 16
social impact