Within the Euromet region a regional key comparison (Euromet.M.P-K1.b) was carried out in order to compare national vacuum standards in the pressure range from 3 × 10-4 Pa to 0.9 Pa. The participants were the BNM-LNE (France), CEM (Spain), IMGC-CNR (Italy), IMT (Slovenia), NPL (United Kingdom), UME (Turkey), and the PTB (Germany) as pilot laboratory. The measurements were carried out from April 2000 to February 2002. Two spinning rotor gauges served as transfer standards and showed a good transport stability. The effective accommodation coefficients of the rotors had to be determined at eight target points at and between 3 × 10-4 Pa and 0.9 Pa. The uncertainty of the generated pressure in the calibration standard was reported as part of the calibration report by each laboratory. All additional uncertainties that were related to the transfer standard were evaluated by the pilot laboratory in order to have a uniform uncertainty analysis for all participants and in order to emphasize the importance of the reported uncertainty of the generated pressure. From the available data a Euromet reference value was calculated at each target pressure. The results from most of the laboratories showed a good agreement with the reference value within the combined uncertainties. A few values of three of the laboratories were significantly off the reference value. At the highest target pressure of 0.9 Pa a linkage to the lowest target pressure at 1 Pa of the key comparison CCM.P-K4 was possible by means of the results of three laboratories that took part in both comparisons.

Results of the regional key comparison Euromet.M.P-K1.b in the pressure range from 3 × 10-4 Pa to 0.9 Pa / Karl, Jousten; Bergoglio, Mercede; Anita, Calcatelli; Jean Noel, Durocher; John, Greenwood; Rifat, Kangi; Jean Claude, Legras; Carmen, Matilla; Janez, Setina. - In: METROLOGIA. - ISSN 1681-7575. - 42:(2005), p. 07001. [10.1088/0026-1394/42/1A/07001]

Results of the regional key comparison Euromet.M.P-K1.b in the pressure range from 3 × 10-4 Pa to 0.9 Pa

BERGOGLIO, MERCEDE;
2005

Abstract

Within the Euromet region a regional key comparison (Euromet.M.P-K1.b) was carried out in order to compare national vacuum standards in the pressure range from 3 × 10-4 Pa to 0.9 Pa. The participants were the BNM-LNE (France), CEM (Spain), IMGC-CNR (Italy), IMT (Slovenia), NPL (United Kingdom), UME (Turkey), and the PTB (Germany) as pilot laboratory. The measurements were carried out from April 2000 to February 2002. Two spinning rotor gauges served as transfer standards and showed a good transport stability. The effective accommodation coefficients of the rotors had to be determined at eight target points at and between 3 × 10-4 Pa and 0.9 Pa. The uncertainty of the generated pressure in the calibration standard was reported as part of the calibration report by each laboratory. All additional uncertainties that were related to the transfer standard were evaluated by the pilot laboratory in order to have a uniform uncertainty analysis for all participants and in order to emphasize the importance of the reported uncertainty of the generated pressure. From the available data a Euromet reference value was calculated at each target pressure. The results from most of the laboratories showed a good agreement with the reference value within the combined uncertainties. A few values of three of the laboratories were significantly off the reference value. At the highest target pressure of 0.9 Pa a linkage to the lowest target pressure at 1 Pa of the key comparison CCM.P-K4 was possible by means of the results of three laboratories that took part in both comparisons.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/31539
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 17
  • ???jsp.display-item.citation.isi??? ND
social impact