An automated temperature-controlled DC Voltage-DC Resistance Multiple Standard (VRMS) for artifact calibration of high end calibrators and multimeters was developed by Measurements International (MI) with the support of the Istituto Nazionale di Ricerca Metrologica (INRiM). The VRMS consists of a 10 V, a 1 Ω, and a 10 kΩ standards selectable via a low thermal scanner. The two resistors are high-stability MI standards while the 10 V standard is based on a low-noise circuit developed by INRIM. A smart feature of the VRMS is its internal algorithm providing updated calibration values of the VRMS standards ensuring their reliably use in the whole period between calibrations. The standards are housed in a thermal box shielding them from temperature changes. The use uncertainties of the VRMS standards are consistent with those for artifact calibration of calibrators and multimeters. The VRMS standards can also act as laboratory references or traveling standards for nterlaboratory comparisons (ILCs). MI is currently commercializing the VRMS as 1330A Artifact Transfer Standard.

An All-In-One Automatic Multiple Standard for Artifact Calibration of High End Electrical Instruments / Galliana, Flavio; Pollarolo, Alessio. - In: CAL LAB. - ISSN 1095-4791. - 32:2(2025), pp. 20-25.

An All-In-One Automatic Multiple Standard for Artifact Calibration of High End Electrical Instruments

FLAVIO GALLIANA
Writing – Original Draft Preparation
;
2025

Abstract

An automated temperature-controlled DC Voltage-DC Resistance Multiple Standard (VRMS) for artifact calibration of high end calibrators and multimeters was developed by Measurements International (MI) with the support of the Istituto Nazionale di Ricerca Metrologica (INRiM). The VRMS consists of a 10 V, a 1 Ω, and a 10 kΩ standards selectable via a low thermal scanner. The two resistors are high-stability MI standards while the 10 V standard is based on a low-noise circuit developed by INRIM. A smart feature of the VRMS is its internal algorithm providing updated calibration values of the VRMS standards ensuring their reliably use in the whole period between calibrations. The standards are housed in a thermal box shielding them from temperature changes. The use uncertainties of the VRMS standards are consistent with those for artifact calibration of calibrators and multimeters. The VRMS standards can also act as laboratory references or traveling standards for nterlaboratory comparisons (ILCs). MI is currently commercializing the VRMS as 1330A Artifact Transfer Standard.
2025
open
File in questo prodotto:
File Dimensione Formato  
CAL_LAB_MAG_2025.pdf

accesso aperto

Tipologia: final published article (publisher’s version)
Licenza: Pubblico - Tutti i diritti riservati
Dimensione 523.11 kB
Formato Adobe PDF
523.11 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/86639
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact