To demonstrate the technical competence of IMBIH with respect to the calibration of air temperature sensors/transducers/meters between -20 °C and 60 °C, in order to get support evidences to claim CMCs in this sector, a hybrid measurement comparison was performed with INRiM which already has published CMCs for this kind of instruments in the range between -20 °C to 80 °C. The comparison involved three UUTs as transfer instrument but only the best one was chosen as the comparison transfer standard. Firstly, the EURAMET TC-T Chair Doc. S. Rudtsch was informed in advance about the comparison and agreed to act as the third independent party then he was substituted by his successor Doc. M. Sadli. The activity was initially planned as part of the EURAMET project P1456 ATM (Air Temperature Metrology) but successively it was planned as an independent exercise in the EURAMET project P1586

L’INRIM e IMBIH hanno organizzato un confronto ibrido bilaterale sulle tarature di strumenti misuratori della temperatura dell’aria tra -20 °C e 60 °C. L’INRIM possiede le CMC in questo settore tra -20 °C e 80 °C, l’IMBIH utilizzerebbe gli eventuali risultati positivi per dimostrare le proprie capacità metrologiche come evidenza per la richiesta delle CMC nel settore in oggetto. Entrambi i laboratori hanno tarato un set di tre termometri e solo il migliore sotto il profilo della stabilità nel tempo è stato scelto come campione di trasferimento. Inizialmente il TC-T Chair Dott. S. Rudtsch ha partecipato all’esercizio come terza parte indipendente, poi è stato sostituito dal suo successore Dott. M. Sadli. L’attività era stata inizialmente pianificata come attività collaterale al progetto EURAMET P1456 ATM (Air Temperature Metrology) ma poi è stata ri-pianificata come attività indipendente nel progetto EURAMET P1586.

Hybrid Comparison between INRIM and IMBIH of air temperature sensors calibrations in the range -20 °C to 60 °C. In the framework of the project EURAMET P1586 / Smorgon, D.; Salerno, R.; Čohodarević, S.. - (2024).

Hybrid Comparison between INRIM and IMBIH of air temperature sensors calibrations in the range -20 °C to 60 °C. In the framework of the project EURAMET P1586

D. Smorgon
Supervision
;
R. Salerno;
2024

Abstract

To demonstrate the technical competence of IMBIH with respect to the calibration of air temperature sensors/transducers/meters between -20 °C and 60 °C, in order to get support evidences to claim CMCs in this sector, a hybrid measurement comparison was performed with INRiM which already has published CMCs for this kind of instruments in the range between -20 °C to 80 °C. The comparison involved three UUTs as transfer instrument but only the best one was chosen as the comparison transfer standard. Firstly, the EURAMET TC-T Chair Doc. S. Rudtsch was informed in advance about the comparison and agreed to act as the third independent party then he was substituted by his successor Doc. M. Sadli. The activity was initially planned as part of the EURAMET project P1456 ATM (Air Temperature Metrology) but successively it was planned as an independent exercise in the EURAMET project P1586
2024
09/2024
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/85820
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact