The measurement of the silicon lattice parameter by a separate-crystal triple-Laue X-ray interferometer is a key step for the realization of the kilogram by counting atoms. Since the measurement accuracy is approaching nine significant digits, a reliable model of the interferometer operation is required to quantify or exclude systematic errors. This paper investigates both analytically and experimentally the effect of the defocus (the difference between the splitter-to-mirror and analyser-to-mirror distances) on the phase of the interference fringes and the measurement of the lattice parameter.
Defocused travelling fringes in a scanning triple-Laue X-ray interferometry setup / Sasso, C P; Mana, G; Massa, E. - In: JOURNAL OF APPLIED CRYSTALLOGRAPHY. - ISSN 0021-8898. - 54:Pt 5(2021), pp. 1403-1408-1408. [10.1107/S1600576721007962]
Defocused travelling fringes in a scanning triple-Laue X-ray interferometry setup
Sasso, C P
;Mana, G;Massa, E
2021
Abstract
The measurement of the silicon lattice parameter by a separate-crystal triple-Laue X-ray interferometer is a key step for the realization of the kilogram by counting atoms. Since the measurement accuracy is approaching nine significant digits, a reliable model of the interferometer operation is required to quantify or exclude systematic errors. This paper investigates both analytically and experimentally the effect of the defocus (the difference between the splitter-to-mirror and analyser-to-mirror distances) on the phase of the interference fringes and the measurement of the lattice parameter.File | Dimensione | Formato | |
---|---|---|---|
vh5143.pdf
accesso aperto
Tipologia:
final published article (publisher’s version)
Licenza:
Creative Commons
Dimensione
706.28 kB
Formato
Adobe PDF
|
706.28 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.