The INRiM static expansion system has been recently modified in order to improve its performance and extend the working range down to 5∙10-4 Pa. A new characterization of the system has been performed and a preliminary estimate of the uncertainty has been evaluated.

Improvement of INRiM static expansion system as vacuum primary standard between 10-4 Pa and 1000 Pa / Astrua, Milena; Mari, Domenico; Pasqualin, Stefano. - Mechanics Quantities / Grandeurs Mécaniques(2019), p. 27007. (Intervento presentato al convegno 19th International Congress of Metrology tenutosi a Paris) [10.1051/metrology/201927007].

Improvement of INRiM static expansion system as vacuum primary standard between 10-4 Pa and 1000 Pa

Astrua, Milena
;
Mari, Domenico;Pasqualin, Stefano
2019

Abstract

The INRiM static expansion system has been recently modified in order to improve its performance and extend the working range down to 5∙10-4 Pa. A new characterization of the system has been performed and a preliminary estimate of the uncertainty has been evaluated.
2019
19th International Congress of Metrology
Paris
open
File in questo prodotto:
File Dimensione Formato  
metrology_cim2019_27007.pdf

accesso aperto

Tipologia: final published article (publisher’s version)
Licenza: Creative Commons
Dimensione 532.9 kB
Formato Adobe PDF
532.9 kB Adobe PDF Visualizza/Apri