A setup for the calibration of lock-in amplifiers in the low frequency range is presented. The setup is based on an uncalibrated voltage source, an ac voltmeter, and calibrated inductive and resistive voltage dividers. The measurand is the relative deviation of the lock-in amplifier reading from the calibrated applied voltage. Examples of calibration of a commercial lock-in amplifier in the 10 μV range, atA setup for the calibration of lock-in amplifiers in the low frequency range is presented. The setup is based on an uncalibrated voltage source, an ac voltmeter, and calibrated inductive and resistive voltage dividers. The measurand is the relative deviation of the lock-in amplifier reading from the calibrated applied voltage. Examples of calibration of a commercial lock-in amplifier in the 10 μV range, at frequencies of 17 Hz and 103 Hz, are given; the calibration uncertainty, relative to the instrument full- scale, is under evaluation and expected to be of the order of 10-4. frequencies of 17 Hz and 103 Hz, are given; the calibration uncertainty, relative to the instrument full- scale, is under evaluation and expected to be of the order of 10 -
Calibration of lock-in amplifiers in the low-frequency range / Cultrera, A.; Tran, N. M. T.; D'Elia, V.; Ortolano, M.; Callegaro, L.. - (2019), pp. 122-125. (Intervento presentato al convegno 23rd IMEKO TC4 International Symposium on Electrical & Electronic Measurements tenutosi a Xi'an, China nel 17-20 September).
Calibration of lock-in amplifiers in the low-frequency range
A. Cultrera
;V. D'Elia;L. Callegaro
2019
Abstract
A setup for the calibration of lock-in amplifiers in the low frequency range is presented. The setup is based on an uncalibrated voltage source, an ac voltmeter, and calibrated inductive and resistive voltage dividers. The measurand is the relative deviation of the lock-in amplifier reading from the calibrated applied voltage. Examples of calibration of a commercial lock-in amplifier in the 10 μV range, atA setup for the calibration of lock-in amplifiers in the low frequency range is presented. The setup is based on an uncalibrated voltage source, an ac voltmeter, and calibrated inductive and resistive voltage dividers. The measurand is the relative deviation of the lock-in amplifier reading from the calibrated applied voltage. Examples of calibration of a commercial lock-in amplifier in the 10 μV range, at frequencies of 17 Hz and 103 Hz, are given; the calibration uncertainty, relative to the instrument full- scale, is under evaluation and expected to be of the order of 10-4. frequencies of 17 Hz and 103 Hz, are given; the calibration uncertainty, relative to the instrument full- scale, is under evaluation and expected to be of the order of 10 -File | Dimensione | Formato | |
---|---|---|---|
IMEKO-TC4-2019_PROCEEDINGS-ISBN.pdf
solo utenti autorizzati
Tipologia:
final published article (publisher’s version)
Licenza:
Non Pubblico - Accesso privato/ristretto
Dimensione
3.24 MB
Formato
Adobe PDF
|
3.24 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.