The CCM.FF-K3.2011 comparison was organized for the purpose of determination of the degree of equivalence of the national standards for air speed over the range 0.5 m/s to 40 m/s. An ultrasonic anemometer and a Laser Doppler anemometer were used as transfer standards. Nine laboratories from three RMOs participated between July 2013 and July 2015 – EURAMET: PTB, Germany; LNE-CETIAT, France; INRIM, Italy; VSL, The Netherlands; E+E, Austria; SIM: NIST, measurements were provided at ambient conditions. All results of independent participants were used in the determination of the key comparison reference value (KCRV) and the uncertainty of the KCRV. The reference value was determined at each air speed separately following “procedure A” presented by M. G. Cox. The degree of equivalence with the KCRV was calculated for each air speed and laboratory. Almost all reported results were consistent with the KCRV.
CCM.FF-K3.2011 Intercomparison for airspeed / Caré, Isabelle; Müller, Harald; Lucas, Peter; Pachinger, Dietmar; Kurihara, Noboru; Lishui, Cui; Su, Chun-Min; Shinder, Iosif; Spazzini, PIER GIORGIO. - online:(2018), pp. 1-12. (Intervento presentato al convegno 10th International Symposium on Fluid Flow Measurement (ISFFM) tenutosi a Queretaro, Mexico nel 21-23 Marzo 2018).
CCM.FF-K3.2011 Intercomparison for airspeed
Pier Giorgio Spazzini
2018
Abstract
The CCM.FF-K3.2011 comparison was organized for the purpose of determination of the degree of equivalence of the national standards for air speed over the range 0.5 m/s to 40 m/s. An ultrasonic anemometer and a Laser Doppler anemometer were used as transfer standards. Nine laboratories from three RMOs participated between July 2013 and July 2015 – EURAMET: PTB, Germany; LNE-CETIAT, France; INRIM, Italy; VSL, The Netherlands; E+E, Austria; SIM: NIST, measurements were provided at ambient conditions. All results of independent participants were used in the determination of the key comparison reference value (KCRV) and the uncertainty of the KCRV. The reference value was determined at each air speed separately following “procedure A” presented by M. G. Cox. The degree of equivalence with the KCRV was calculated for each air speed and laboratory. Almost all reported results were consistent with the KCRV.File | Dimensione | Formato | |
---|---|---|---|
K3_2011paper.pdf
accesso aperto
Descrizione: Articolo Principale
Tipologia:
final published article (publisher’s version)
Licenza:
Pubblico - Tutti i diritti riservati
Dimensione
992.32 kB
Formato
Adobe PDF
|
992.32 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.