Current transformers (CTs) are precision devices that scale high currents down to values that can be easily handled by measurement equipment. To support CT applications in revenue metering, a comparison on AC current transformer calibration systems was performed among 15 European national metrology institutes using a precision CT as the travelling device. The first comparison results for the transformation ratios (4, 5, 6, 8, 10) kA/5 A of the travelling CT at nominal burden of 5 VA and 15 VA indicate good agreement between the participating laboratories. The main differences are found for phase displacement, at least partly caused by the instability of the traveling standard.

International comparison of current transformer calibration systems up to 10 kA at 50 Hz frequency / Draxler, K.; Styblikova, R.; Rietveld, G.; van den Brom, H.; Schnaitt, M.; Waldmann, W.; Dimitrov, E.; Cincar Vujovic, T.; Paczek, B.; Sadkowski, G.; Crotti, Gabriella; Martin, R.; Garnacho, F.; Blanc, I.; Kampfer, R.; Mester, C.; Wheaton, A.; Mohns, E.; Bergman, A.; Hammarquist, M.; Cayci, H.; Hallstrom, J.; Suomalainen, E. P.. - (2016), pp. 1-2. (Intervento presentato al convegno 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016)) [10.1109/CPEM.2016.7540493].

International comparison of current transformer calibration systems up to 10 kA at 50 Hz frequency

CROTTI, GABRIELLA;
2016

Abstract

Current transformers (CTs) are precision devices that scale high currents down to values that can be easily handled by measurement equipment. To support CT applications in revenue metering, a comparison on AC current transformer calibration systems was performed among 15 European national metrology institutes using a precision CT as the travelling device. The first comparison results for the transformation ratios (4, 5, 6, 8, 10) kA/5 A of the travelling CT at nominal burden of 5 VA and 15 VA indicate good agreement between the participating laboratories. The main differences are found for phase displacement, at least partly caused by the instability of the traveling standard.
2016
2016 Conference on Precision Electromagnetic Measurements (CPEM 2016)
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/54713
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 0
social impact