A temperature controlled 1 –10 kΩ Standard resistors setup was developed at National Institute of Metrological Research, (INRIM). The aim of this realization was the involvement of the setup resistors in the traceability transfer process to high accuracy multifunction electrical instruments used in Secondary Electrical Calibration Laboratories or even their use as primary Standards in high level Laboratories or Institutes. The 1 –10 kΩ Standards are formed respectively by two 10 and 100 k parallel connected resistors nets inserted in a temperature controlled aluminium box. Construction details, temperature and power coefficients, stability data and preliminary mid-term use uncertainty budgets of the two setup Standards are given. Their first short time (2 h) stability were on the order of few parts of 10–8. A test to calibrate a multifunction calibrator gave satisfactory results
1 Ω–10 kΩ high precision Resistance Setup to calibrate multifunction electrical instruments / Capra, PIER PAOLO; Galliana, Flavio; M., Lanzillotti. - (2014), pp. 945-949. (Intervento presentato al convegno 20th IMEKO TC-4 International Symposium Measurement of Electrical Quantities"Research on Electrical and Electronic Measurement for the Economic Upturn" 18th TC-4 Workshop on ADC and DAC Modelling and Testing tenutosi a BENEVENTO ITALY nel September 15 - 17, 2014).
1 Ω–10 kΩ high precision Resistance Setup to calibrate multifunction electrical instruments
CAPRA, PIER PAOLO;GALLIANA, FLAVIO;
2014
Abstract
A temperature controlled 1 –10 kΩ Standard resistors setup was developed at National Institute of Metrological Research, (INRIM). The aim of this realization was the involvement of the setup resistors in the traceability transfer process to high accuracy multifunction electrical instruments used in Secondary Electrical Calibration Laboratories or even their use as primary Standards in high level Laboratories or Institutes. The 1 –10 kΩ Standards are formed respectively by two 10 and 100 k parallel connected resistors nets inserted in a temperature controlled aluminium box. Construction details, temperature and power coefficients, stability data and preliminary mid-term use uncertainty budgets of the two setup Standards are given. Their first short time (2 h) stability were on the order of few parts of 10–8. A test to calibrate a multifunction calibrator gave satisfactory resultsI documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.