A comparison of a Pt/Pd calibration above the Ag point between the INRIM and NMC was arranged with the aims of evaluating measurement systems and exploiting the potential of the Pt/Pd thermocouples. Two commercial Pt/Pd thermocouples were used as transfer thermometers. A calibration method using a blackbody cavity as a transfer source and a radiation thermometer as a reference thermometer was adopted in both institutes. The T90 carried by the radiation thermometers is established by an extrapolation technique for INRIM and by scale realization according to ITS-90 definition for NMC and, therefore, this exercise is also a useful comparison of different approaches to disseminate T90 above the Ag point. The comparison results are presented and analyzed.
INRIM–NMC Comparison of Pt/Pd Calibration Above the Ag Point / Battuello, Mauro; Girard, Ferruccio; Wang, L.. - In: INTERNATIONAL JOURNAL OF THERMOPHYSICS. - ISSN 0195-928X. - 31:(2010), pp. 1444-1455. [10.1007/s10765-010-0805-x]
INRIM–NMC Comparison of Pt/Pd Calibration Above the Ag Point
BATTUELLO, MAURO;GIRARD, FERRUCCIO;
2010
Abstract
A comparison of a Pt/Pd calibration above the Ag point between the INRIM and NMC was arranged with the aims of evaluating measurement systems and exploiting the potential of the Pt/Pd thermocouples. Two commercial Pt/Pd thermocouples were used as transfer thermometers. A calibration method using a blackbody cavity as a transfer source and a radiation thermometer as a reference thermometer was adopted in both institutes. The T90 carried by the radiation thermometers is established by an extrapolation technique for INRIM and by scale realization according to ITS-90 definition for NMC and, therefore, this exercise is also a useful comparison of different approaches to disseminate T90 above the Ag point. The comparison results are presented and analyzed.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.