This paper describes the scattering parameters magnitude measurement and uncertainty comparison using different vector network analyzer (VNA) calibration algorithms at millimeter frequencies with WR10 (75 to 110 GHz) waveguides. The state of the art two port calibration algorithms Thru-Reflect- Line (TRL) and Quick Short-Open-Load-Thru (QSOLT) are considered in this comparison. The dimensional measurements for Shim WR10 are also carried out.
Error propagation with different VNA calibration techniques at millimeter frequencies / Sellone, Marco; N., Shoaib; A., Ferrero; Oberto, Luca; Brunetti, Luciano. - (2013), pp. 24.1-24.3. (Intervento presentato al convegno 82nd ARFTG Microwave Measurement Conference tenutosi a Columbus, Ohio, USA nel 18-21/11/2013) [10.1109/ARFTG-2.2013.6737351].
Error propagation with different VNA calibration techniques at millimeter frequencies
SELLONE, MARCO;OBERTO, LUCA;BRUNETTI, LUCIANO
2013
Abstract
This paper describes the scattering parameters magnitude measurement and uncertainty comparison using different vector network analyzer (VNA) calibration algorithms at millimeter frequencies with WR10 (75 to 110 GHz) waveguides. The state of the art two port calibration algorithms Thru-Reflect- Line (TRL) and Quick Short-Open-Load-Thru (QSOLT) are considered in this comparison. The dimensional measurements for Shim WR10 are also carried out.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.