The results of an international comparison of absolute gravimeters held in Walferdange Luxembourg in November 2003 are presented. The absolute meters agreed with one another with a standard deviation of less than 2 µGal (1 Gal = 1 cm/s2) (if we exclude one prototype instrument from the analysis). For the first time, the ability of the operators was put to the test. The comparison indicates that the errors due to the operator are less than 1 µGal, i.e. within the observational errors.
Results of the International Comparison of Absolute Gravimeters in Walferdange (Luxembourg) of November 2003 / Francis, O; VAN DAM, T; Amalvict, M; DE ANDRADE SOUSA, M; Bilker, M; Billson, R; D'Agostino, Giancarlo; Desogus, S; Falk, R; Germak, ALESSANDRO FRANCO LIDIA; Gitlein, O; Jonhson, D; Klopping, F; Kostelecky, J; Luck, B; Mäkinen, J; Mclaughlin, D; Nunez, E; Origlia, C; Palinkas, V; Richard, P; Rodriguez, E; Ruess, D; Schmerge, D; Thies, S; Timmen, L; VAN CAMP, M; VAN WESTRUM, D; Wilmes, H.. - 129:(2005), pp. 272-275. (Intervento presentato al convegno Gravity, Geoid and Space Missions, GGSM 2004, IAG International Symposium tenutosi a Porto, Portugal nel 30 August - 3 September 2004) [10.1007/3-540-26932-0_47].
Results of the International Comparison of Absolute Gravimeters in Walferdange (Luxembourg) of November 2003
D'AGOSTINO, GIANCARLO;GERMAK, ALESSANDRO FRANCO LIDIA;ORIGLIA C;
2005
Abstract
The results of an international comparison of absolute gravimeters held in Walferdange Luxembourg in November 2003 are presented. The absolute meters agreed with one another with a standard deviation of less than 2 µGal (1 Gal = 1 cm/s2) (if we exclude one prototype instrument from the analysis). For the first time, the ability of the operators was put to the test. The comparison indicates that the errors due to the operator are less than 1 µGal, i.e. within the observational errors.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.