Electrostatic discharge (ESD) immunity tests are widely carried out in order to assess the immunity level of electrical and electronic equipment. ESD is a severe source of interference which can produce damage, upset or failures in any electronic system. A careful calibration procedure is required to assure that the ESD generator parameters fulfill the standard specifications. In this paper, the main critical aspects involved in the calibration measurement setup are dealt with and a simple modeling support is given. On the basis of the obtained results, some useful conclusions and practical suggestions are finally drawn.
Critical Aspects in Calibration of ESD Generators / Morando, A; Borsero, Michele; Sardi, Angelo; Vizio, G.. - In: MEASUREMENT SCIENCE REVIEW. - ISSN 1335-8871. - 11:(2011), pp. 23-28. [10.2478/v10048-011-0002-5]
Critical Aspects in Calibration of ESD Generators
BORSERO, MICHELE;SARDI, ANGELO;
2011
Abstract
Electrostatic discharge (ESD) immunity tests are widely carried out in order to assess the immunity level of electrical and electronic equipment. ESD is a severe source of interference which can produce damage, upset or failures in any electronic system. A careful calibration procedure is required to assure that the ESD generator parameters fulfill the standard specifications. In this paper, the main critical aspects involved in the calibration measurement setup are dealt with and a simple modeling support is given. On the basis of the obtained results, some useful conclusions and practical suggestions are finally drawn.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.