The capacitance-charging method is a well-established and handy technique for the generation of dc current in the 100 pA rangeor lower. The method involves a capacitance standard and a sampling voltmeter, highly stable devices easy to calibrate; it is robust and insensitive to the voltage burden of the instrument being calibrated. We propose here a range extender amplifier, which can be employed as a plug-in component in existing calibration setups, and allows the generation of currents in the 1 nA range. The extender has been employed in the INRIM setup and validated with two comparisons at 100 pA and 1 nA current levels. The calibration accuracy achieved on a top-class instrument is 5E-5 at 1 nA.
Generation of reference dc currents at 1 nA level with the capacitance-charging method / Callegaro, Luca; Capra, PIER PAOLO; D'Elia, V.; Galliana, Flavio. - In: IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT. - ISSN 0018-9456. - 63:7(2014), pp. 1779-1782. [http://www.inrim.it/Callegaro2014_1nA.pdf]
Generation of reference dc currents at 1 nA level with the capacitance-charging method
CALLEGARO, LUCA;CAPRA, PIER PAOLO;GALLIANA, FLAVIO
2014
Abstract
The capacitance-charging method is a well-established and handy technique for the generation of dc current in the 100 pA rangeor lower. The method involves a capacitance standard and a sampling voltmeter, highly stable devices easy to calibrate; it is robust and insensitive to the voltage burden of the instrument being calibrated. We propose here a range extender amplifier, which can be employed as a plug-in component in existing calibration setups, and allows the generation of currents in the 1 nA range. The extender has been employed in the INRIM setup and validated with two comparisons at 100 pA and 1 nA current levels. The calibration accuracy achieved on a top-class instrument is 5E-5 at 1 nA.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.