INRIM's gas flow standard is realized by means of three facilities (MicroGas, BellGas and MeGas) The facilities must be periodically calibrated in order to guarantee and update the measurements capabilities and therefore the uncertainties associated with the standard. This Technical Report concerns the second calibration of the MeGas facility, detailing the calculation of the uncertainty associated with the diameter and the displacement of the MeGas piston. The Technical Report INRIM 15/2019 describes the procedure of the calibration.
Analisi delle incertezze associate al campione nazionale di portate di gas dell’INRIM denominato MeGas / Bisi, Marco; Piccato, Aline; Bertiglia, Fabio; Spazzini, PIER GIORGIO; LA PIANA, Gaetano; Bellotti, Roberto. - (2020).
Analisi delle incertezze associate al campione nazionale di portate di gas dell’INRIM denominato MeGas
Marco Bisi;Aline Piccato;Fabio Bertiglia;Pier Giorgio Spazzini;Gaetano La Piana;Roberto Bellotti
2020
Abstract
INRIM's gas flow standard is realized by means of three facilities (MicroGas, BellGas and MeGas) The facilities must be periodically calibrated in order to guarantee and update the measurements capabilities and therefore the uncertainties associated with the standard. This Technical Report concerns the second calibration of the MeGas facility, detailing the calculation of the uncertainty associated with the diameter and the displacement of the MeGas piston. The Technical Report INRIM 15/2019 describes the procedure of the calibration.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.