The inter-comparison has been organised within EURAMET with project number 1268 [1] and has been also registered in the BIPM KCDB as EURAMET.T-S3 [2]. The inter-comparison was approved during the 2013 EURAMET Technical Committee on Thermometry (TC-T) meeting and officially started in October 2013, the final measurements were performed in November 2016 and the last participant’s reports were received in March 2018. The report describes the measurements results obtained by INRIM in the calibration of two Pt/Pt thermocouples from 420 °C to 1492 °C both at fixed points and by comparison. This inter-comparison is intended to be used to support the calibration and measurement capabilities (CMCs) of the participants in the calibration of thermocouples. INRIM measurements were performed at INRIM laboratory of primary thermometry, industrial thermometry and radiation thermometry and results of the calibration carried out with different methods are compared.

The EURAMET Project N° 1268 “Calibration of thermocouples from 419,527 ºC (freezing point of Zn) up to 1492 ºC (melting point of the Pd-C eutectic), by the temperature fixed points and comparison methods” (EURAMET.T-S3) / Bertiglia, F.; Braccialarghe, G.; Dematteis, R.; Florio, M.; Girard, F.; Iacomini, L.; Lopardo, G.; Steur, P. P. M.. - (2021).

The EURAMET Project N° 1268 “Calibration of thermocouples from 419,527 ºC (freezing point of Zn) up to 1492 ºC (melting point of the Pd-C eutectic), by the temperature fixed points and comparison methods” (EURAMET.T-S3)

F. Bertiglia;G. Braccialarghe;R. Dematteis;M. Florio;F. Girard;L. Iacomini;G. Lopardo;P. P. M. Steur
2021

Abstract

The inter-comparison has been organised within EURAMET with project number 1268 [1] and has been also registered in the BIPM KCDB as EURAMET.T-S3 [2]. The inter-comparison was approved during the 2013 EURAMET Technical Committee on Thermometry (TC-T) meeting and officially started in October 2013, the final measurements were performed in November 2016 and the last participant’s reports were received in March 2018. The report describes the measurements results obtained by INRIM in the calibration of two Pt/Pt thermocouples from 420 °C to 1492 °C both at fixed points and by comparison. This inter-comparison is intended to be used to support the calibration and measurement capabilities (CMCs) of the participants in the calibration of thermocouples. INRIM measurements were performed at INRIM laboratory of primary thermometry, industrial thermometry and radiation thermometry and results of the calibration carried out with different methods are compared.
2021
4
none
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11696/85119
 Attenzione

Attenzione! I dati visualizzati non sono stati sottoposti a validazione da parte dell'ateneo

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact