In 1995 the 1000 kN deadweight force standard machine was designed and installed in the Bray Pavilion at INRiM. The machine is an instrument suitable for generating known forces and is used as a reference standard for the calibration or verification of force transducers and load cells. Despite the positive international comparisons, it was necessary to carry out a new verification of the MCF 1000 to evaluate a possible natural deterioration of the machine. From October 2018 to June of 2019 measurements and analysis of the associated uncertainties were carried out. Results are within the limits of the declared CMC.
Nel 1995 è stata progettata e installata nel Padiglione Bray presso INRiM la macchina campione di forza a pesi diretti MCF da 1000 kN (MCF 1000). La macchina in questione è uno strumento atto alla generazione di forze note ed è utilizzata come campione di riferimento per la taratura o la verifica di trasduttori di forza e celle di carico. Nonostante i positivi confronti internazionali intercorsi, si è ritenuto necessario provvedere a una verifica della MCF 1000 al fine di valutare un eventuale deperimento naturale della macchina. Tra ottobre 2018 e giugno 2019 sono state effettuate le misure e l’analisi delle incertezze associate. I risultati ottenuti rientrano nei limiti della CMC dichiarata.
Autotaratura della macchina campione di forza a pesi diretti da 1 MN “MCF 1000” presso INRiM / Prato, A.; Mazzoleni, F.; Facello, A.; Germak, A.. - 04/2020:(2020), pp. 1-35. [10.13140/RG.2.2.23728.79369]
Autotaratura della macchina campione di forza a pesi diretti da 1 MN “MCF 1000” presso INRiM
A. PratoWriting – Original Draft Preparation
;F. MazzoleniData Curation
;A. FacelloFormal Analysis
;A. Germak
Writing – Review & Editing
2020
Abstract
In 1995 the 1000 kN deadweight force standard machine was designed and installed in the Bray Pavilion at INRiM. The machine is an instrument suitable for generating known forces and is used as a reference standard for the calibration or verification of force transducers and load cells. Despite the positive international comparisons, it was necessary to carry out a new verification of the MCF 1000 to evaluate a possible natural deterioration of the machine. From October 2018 to June of 2019 measurements and analysis of the associated uncertainties were carried out. Results are within the limits of the declared CMC.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.