In this technical report we will describe the project and construc- tion of the new Calorimetric Thermal Voltage Converter which will be used in the INRiM High Frequency Laboratory to extend the metro- logical capabilities in the power measurements field. In fact the mi- crocalorimetric technique used until now is limited in accuracy below 10 MHz, because, under this frequency value, the noise covers the sig- nals to be detected. With a CTVC we will measure an HF Power through an RF Voltage measurement, using higher power levels which will be more efficiently measurable.
Design and realization of a Calorimetric Thermal Voltage Converter for frequencies up to 1 GHz / Sellone, Marco; Capelli, A.. - 50:(2007), pp. 1-33.
Design and realization of a Calorimetric Thermal Voltage Converter for frequencies up to 1 GHz
SELLONE, MARCO;
2007
Abstract
In this technical report we will describe the project and construc- tion of the new Calorimetric Thermal Voltage Converter which will be used in the INRiM High Frequency Laboratory to extend the metro- logical capabilities in the power measurements field. In fact the mi- crocalorimetric technique used until now is limited in accuracy below 10 MHz, because, under this frequency value, the noise covers the sig- nals to be detected. With a CTVC we will measure an HF Power through an RF Voltage measurement, using higher power levels which will be more efficiently measurable.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.